TFG-3058
ONLINE FLUORIDE ANALYZER
TFG-3058 is automatic and oline fluoride analyzer ,working principle :The fluoride ion in the water sample reacts with the fluorine reagent and cerium nitrate in an acetate buffer solution to form a colored complex, and the analyzer detects the change in the color and converts the change into a fluoride ion value. The amount of colored complex formed is equivalent to the amount of fluoride ion.mainly application for driking water, surface water etc.
Standard
Fluorine reagent spectrophotometry
Measuring range
0.00-15.00 mg/L,0.5~50mg/L
Accuracy
±0.05mg/L when measured value ≤0.5mg/L; ±10% when measured value >0.05mg/L
Repeatability
≤±5%
Stability
≤±10% in 24 hours
Measurement period
The minimum measurement period is 50 minutes, according to the actual water sample, The digestion time can be modified from 5 to 120 minutes.
Sampling period
Time interval(55~9999min adjustable by yourself) and Integral time measurement mode;
Calibration period
Adjustable in 1~99days;
Maintenance period
Usually one time per month and 30 minutes each time
Output
4~20mA( 2 ways) RS232,RS485
It should be indoor and temperature can be adjustable. The recommend temperature is +5~28℃;humidity≤90%(no dew)
Power
AC220±10% V,50±10% Hz,5 A;
Dimension size
Height 1450×Width510×Length 450 mm;
Data saved
The data can be saved when abnormal alarms and power off.
Display
Touch screen and Command Input
Other
When Reset after abnormal alarm and power on after off, the instrument automatically discharges residual reactants in the instrument and automatically resumes working status.